The HI-98190 combines all of the features of a benchtop into a portable IP67 rated waterproof enclosure making them one of the most versatile meters on the market.
Designed for universal applications, HI-98190 measures pH/ORP and temperature. Exchange out the pH probe for an ORP probe to obtain mV readings in the ±2000 mV range.
A backlit, graphic LCD provides easy to read resolution even in low-lit areas. A combination of dedicated and soft keys allows for intuitive operation in a choice of languages while the extended battery life assures long operation.
Quick Connect Probe
The HI-98190 features the HI-12963 titanium bodied pH/temperature electrode with a quick connect DIN connector to make attaching and removing the probe simple and easy.
Data Logging
The log-on demand feature allows users to store up to 200 sample that can be later transferred to a PC with the HI-920015 USB cable and HI-92000 software.
Key Features
- IP67 rated waterproof, rugged enclosure
- CAL Check™ - alerts users to problems during calibration including dirty/broken electrodes, contaminated buffer and overall probe condition.
- Automatic or manual temperature compensation - pH sensors incorporate a built-in temperature sensor.
- Calibration - up to a five-point calibration with seven standard buffers and five custom buffers available.
- Intuitive keypad - most of the available options such as GLP information, help, range, calibration and backlight have a dedicated button
- Approximately 200 hour battery life powered by (4) 1.5V AA batteries
Supplied with
- HI-12963 titanium body pH electrode
- HI-7004M
- HI-7007M
- 2 x sachets of electrode cleaning solution
- Rugged carrying case
- 100mL plastic beaker (2)
- HI-92000 PC software
- HI-920015 micro USB cable
- Batteries
- Instruction manual
- Quality certificate
| pH - Range |
|
-2.0 to 20.0 pH; -2.00 to 20.00 pH; -2.000 to 20.000 pH |
| Resolution |
|
0.1 pH; 0.01 pH; 0.001 pH |
| Accuracy |
|
±0.1 pH; ±0.002 pH |
| Calibration |
|
up to five-point calibration, seven standard buffers available (1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45) and five custom buffers |
| Temperature Compensation |
|
automatic or manual from -20.0 to 120.0°C |
| mV - Range |
|
±2000 mV |
| Resolution |
|
0.1 mV |
| Accuracy |
|
±0.2 mV |
| Relative mV Offset Range |
|
±2000 mV |
| Temperature - Range |
|
-20.0 to 120.0°C |
| Resolution |
|
0.1°C |
| Accuracy |
|
±0.4°C |
| Additional Specifications: |
|
|
| pH Probe |
|
HI-12963 titanium body, pH electrode with internal temperature sensor, quick DIN connector and 1m cable |
| Slope Calibration |
|
from 80 to 110% |
| Log-on-demand |
|
200 samples (100 each pH/mV range) |
| PC Connection |
|
opto-isolated USB with HI-92000 software and micro USB cable |
| Input Impedance |
|
10¹²Ω |
| Battery Type/Life |
|
1.5 C AA batteries (4) / approximately 200 hours of continuous use without backlight (50 hours with backlight) |
| Auto-off |
|
user selectable; 5, 10, 30, 60 min, disabled |
| Environment |
|
0 to 50°C; RH 100 % IP67 |
| Dimensions/Weight |
|
185 x 93 x 35.2 mm / 400g |